Increase Font Size Decrease Font Size

Safeguards Analytical Laboratory

Quality Assurance and Computer Services group

Warning and Control Limits for SAL control measurements

(limits for the bias of a single QC measurement result)

Method

control material

measurand

limits

     

control

warning

Gamma-Spec NBS020, NBS030 235U wt%

0.50

0.33

HRGS NBS947, SAL9984 241Am

1.97

1.47

    238:239

3.56

2.66

    240:239

1.93

1.46

    241:239

1.50

1.13

Alpha-Spec NBS947, AEA90099 act.ratio

1.17

0.85

IDMS LSD LSD control Pu conc.

0.41

0.31

    U conc

0.22

0.17

IDMS product UK82522 or mixture Pu conc.

0.48

0.36

  NBL112A (NBS960) or mixture U conc

0.43

0.30

MassSpec Pu NBS947 240:239

0.19

0.14

    242:239

0.51

0.38

  UK-Pu-3 240:239

0.11

0.085

    242:239

0.21

0.16

  UK-Pu-4 240:239

0.18

0.13

  UK-Pu-5 240:239

0.18

0.13

    242:239

0.32

0.23

    244:239

0.36

0.27

  UK-Pu-6 240:239

0.15

0.11

  NBL-128 242:239

0.20

0.15

  CBNM-047a 244:239

0.30

0.22

MassSpec U NBS010 235:238

0.36

0.27

  NBS020a, NBS030a 235:238

0.20

0.15

  NBS100 to NBS930

NBL116, GUS3568, CBNM-072/2

235:238

0.22

0.16

  CBNM-072/2, GUS3568 233:238

0.26

0.20

  GUS3568 236:238

0.15

0.11

McD&S UK-Pu-1 (as control material) Pu conc.

0.23

0.17

D&G Pu-Lab EC-110 U conc.

0.26

0.19

D&G U-Lab EC-110 U conc.

0.12

0.09

These Warning and Control Limits consider the uncertainty of the certified reference values and the uncertainty of physical constants like radioactive decay half-lifes. Warning and Control Limits are defined in accordance with ISO Standard 3534 "Statistics -- Vocabulary and symbols